Letters

Inappropriate implantable defibrillator discharges from lead failure

Chin V Hiew and James W Leitch
MJA 2009; 190 (2): 97-98

To the Editor: A 77-year-old woman with a history of idiopathic long-QT syndrome (presumed genetic) presented to hospital with a storm of defibrillator shocks from her implantable cardioverter defibrillator (ICD). She had a single-chamber Guidant Ventak Prizm VR, model 1850, and a Guidant ventricular lead, model 0148 (Guidant Corporation, Indianapolis, Ind, USA) implanted 12 months earlier after an episode of syncope. The lead had been implanted by left subclavian venous access. The patient’s postoperative course and responses to routine questioning indicated that the device was functioning normally.

After presenting with device discharges, a 12-lead electrocardiogram showed oversensing suggestive of lead dysfunction. Stored electrograms from the defibrillator showed artifactual signals that were typical of lead disruption. These signals had triggered inappropriate detection of ventricular arrhythmia and subsequent device shocks (Box 1). The pacing threshold and lead impedance were normal. On close examination of the patient’s chest x-ray, a line of radiolucency was detected within the lead between the first rib and clavicle. At reoperation, the lead was extracted with minimum difficulty and the site of the insulation break was detected (Box 2). A new lead was implanted and the patient was discharged home well. Subsequent follow-up was uneventful.

Lead failure can occur any time after implantation of an ICD, and should be suspected when patients present with inappropriate ICD discharges.1 This case illustrates oversensing as a result of the lead’s insulation being disrupted by crush compression between the first rib and clavicle. Alternative implantation techniques, such as cephalic vein cut-down or extrathoracic axillary vein puncture, have been suggested for venous access to avoid this complication.2

1 Stored electrograms from the patient’s implantable cardioverter defibrillator showed artifactual signals had triggered inappropriate detection of ventricular arrhythmia and subsequent device shocks (arrow)

2 Implantable cardioverter defibrillator lead after extraction showing the break in the insulation (arrow)

Chin V Hiew, Interventional Cardiology FellowJames W Leitch, Cardiologist and Electrophysiologist

Cardiovascular Department, John Hunter Hospital, Newcastle, NSW.

cvhiewAThotmail.com

  1. Roelke M, O’Nunain SS, Osswald S, et al. Subclavian crush syndrome complicating transvenous cardioverter defibrillator systems. Pacing Clin Electrophysiol 1995; 18: 973-980. <PubMed>
  2. Belott P. How to access the axillary vein. Heart Rhythm 2006; 3: 366-369. <PubMed>

(Received 20 Mar 2008, accepted 10 Jun 2008)